GC Inlet Liners: Wool or None, Taper or Straight, and Matching One to the Injection
In one liner evaluation, endrin breakdown was 1.2 % on injection one and 12.2 % on injection 101. Nothing failed, and the number climbed anyway.
Equipment Hub / Chromatography & sample prep
Inlet liners, septa, vials, caps, syringe filters, filter paper, pipette tips and cuvettes: the small parts that decide data quality.
Parts covered: GC inlet liners and septa, autosampler vials and caps, syringe filters, pipette tips, cuvettes.
Choose the right part
In one liner evaluation, endrin breakdown was 1.2 % on injection one and 12.2 % on injection 101. Nothing failed, and the number climbed anyway.
Grade 5 and Grade 42 are both rated 2.5 µm. Their listed filtration times are 1,420 and 1,870 seconds, and neither listing names the method behind the number.
At 280 nm, UV-grade quartz passes 88% of the light and optical glass passes 5%. The band that nucleic acid work depends on sits on the wrong side of that cliff.
Over 56 hours, no cap tested held a volatile analyte. The best lost 26%, the worst 68% — and the difference was the injector, not the cap.
Through 0.20 µm nylon the average analyte loss in one 2022 study was 71%, while a cannabinoid note in methanol put nylon among the best. Both can be right.
A 250 °C inlet set point can leave the septum 100 °C cooler, or nearly as hot on an older instrument. That difference decides which disc you need.
In one manufacturer study of fifteen tip suppliers, thirteen broke the permissible error at least once, and every one of the failing tips still fitted the pipette.