ULT Freezers: -70 °C against -80 °C, 20 to 40 Percent Energy Savings by Source, and 41 against 44 Hours to 0 °C
Moving the setpoint 10 K warmer saves somewhere between a fifth and two fifths of the energy, depending on whom you ask, and costs three hours of rescue time after a power cut.

Disclosure: We may earn a commission if you purchase through links on this page, at no additional cost to you. Our selection criteria do not depend on commission; see our affiliate disclosure.
Published figures for moving a ULT freezer from −80 °C to −70 °C: about 20 to 22 percent less energy at Copenhagen across four new freezers, 28.6 percent at Edinburgh with VWR on three units, 25 to 30 percent at Bath, 37 percent in a California study quoted by CU Boulder, and up to 40 percent from the US Department of Energy as quoted by a distributor. Warm-up to 0 °C after a power loss: 44 hours at −80 °C against 41 hours at −70 °C at Bath.
Energy saving from −70 °C, by source
- University of Copenhagen, plug load of four new freezers: total consumption 20 to 22 percent lower at −70 °C.
- University of Edinburgh with VWR, July 2015, three New Brunswick U570 units: 6.0 kWh/day at −70 °C against 8.4 kWh/day at −80 °C, a saving of 28.6 percent.
- University of Bath guide: roughly 25 to 30 percent.
- Dutch Green Labs, citing Graham et al., 2024: about 28 percent per freezer.
- CU Boulder, quoting a 2016 California study: an average of 37 percent for standard-efficiency and energy-efficient units.
- US Department of Energy, as quoted by a distributor: up to 40 percent, with 20 to 40 percent given as the range in appropriate applications.
- A University of British Columbia slide deck: 30 to 40 percent, no study named.
The range across these is 20 to 40 percent, a factor of 2 between the lowest and highest. Edinburgh's three units were three years old in a room held at 17 to 18 °C; Copenhagen's were new. None of the sources says how much of the gap comes from freezer age, load or room temperature.
The Edinburgh set point table
| Set point | kWh/day | Against −80 °C |
|---|---|---|
| −60 °C | 4.9 | 41.7 % lower |
| −70 °C | 6.0 | 28.6 % lower |
| −75 °C | 7.1 | 15.5 % lower |
| −80 °C | 8.4 | reference |
| −85 °C | 9.6 | 14.3 % higher |
A year later the same report gives 8.5 kWh/day at −80 °C and a 30 percent saving at −70 °C.
Arithmetic on that table: −80 to −70 °C is 2.4 kWh/day, or 876 kWh a year per freezer. Per degree, the cost is 0.24 kWh/day between −80 and −85 °C, 0.26 between −75 and −80, 0.22 between −70 and −75, and 0.11 between −60 and −70. The marginal cost of a degree is therefore about 0.22 to 0.26 kWh/day across −70 to −85 °C, and roughly half that below −70 °C.
Manufacturer comparison sheet
A manufacturer's comparison sheet lists three models at both set points: 4.5 kWh/day at −70 °C against 5.4 at −80 °C; 5.9 against 8.2; and 6.2 against 7.9. The savings work out at 16.7, 28.0 and 21.5 percent, two of the three below Edinburgh's 28.6 percent. It is a data sheet, so it is used here only as one more spread of values.
ENERGY STAR numbers
Under ENERGY STAR Laboratory Grade Refrigerators and Freezers Version 2.0, a ULT freezer is one capable of holding set points between −70 °C and −80 °C, energy is tested at −75 °C, and the maximum daily consumption is 0.46 kWh/day/ft³ below 20 ft³ and 0.35 kWh/day/ft³ at 20 ft³ or more, effective 30 June 2025 by date of manufacture.
An older summary of the previous programme gives 0.55 kWh/day/ft³ or less at −75 °C, and average values from California test data: 0.45 at −80 °C, 0.40 at −75 °C and 0.32 at −70 °C for high-efficiency models, against 0.84 at −80 °C for standard models.
Two listed products: a 24.09 ft³ model at 0.31 kWh/day/ft³, which is 7.5 kWh/day, and a 25.71 ft³ model at 0.36 kWh/day/ft³, which is 9.3 kWh/day. For a 24.09 ft³ unit the Version 2.0 limit allows 0.35 × 24.09, or 8.4 kWh/day; the older 0.55 figure allowed 13.2 kWh/day for the same size. The second listing at 0.36 sits just above the new limit and would fail it, though it is not established whether that unit was certified under the old version or the new one.
Absolute consumption: two different ranges
At −80 °C the sources give 8.4 kWh/day for 2015-vintage units, 5.4 to 8.2 from the comparison sheet, up to 20 kWh/day in one sustainability newsletter, and 16 to 22 kWh/day in a UK laboratory medicine document that contrasts it with 8 to 10 kWh/day for a typical home.
At 20 kWh/day a freezer uses 7,300 kWh a year; at 8.4 kWh/day it uses 3,066. The ratio is 2.4, and no source explains it. The likeliest reading is that the older and newer figures describe different generations of equipment, but that is an inference rather than a published statement.
Power failure: how long to warm
Edinburgh, with water and ice in polystyrene boxes across three shelves:
| Target | −70 °C start, bottom / middle / top | −80 °C start, bottom / middle / top |
|---|---|---|
| −50 °C | 3h25 / 5h25 / 4h25 | 4h50 / 5h50 / 5h00 |
| −20 °C | 14h50 / 19h10 / 18h10 | 16h25 / 19h45 / 18h25 |
| 0 °C | 36h55 / 43h50 / 42h15 | 39h40 / 46h50 / 45h25 |
The −80 minus −70 differences are 1h25, 0h25 and 0h35 to −50 °C; 1h35, 0h35 and 0h15 to −20 °C; and 2h45, 3h00 and 3h10 to 0 °C. Bath gives 41 hours at −70 °C against 44 hours at −80 °C to reach 0 °C, and 23 hours from −60 °C; the three-hour gap is 6.8 percent of 44 hours.
Two other published numbers do not match. A slide deck citing Faugeroux (UC Riverside, 2016) gives about nine hours from −80 °C to −50 °C for most modern freezers, against Edinburgh's 4h50 to 5h50 for the same step — almost double. Loading, model and door seal condition all differ and the sources do not give enough to say which dominates. The same slides estimate that a −70 °C setting loses 1.5 to 3 hours of rescue time and leaves at least 6 hours for intervention.
What the sample evidence shows
A University of Virginia comparison reported in a slide deck, one year at each temperature with three animals per condition:
| Tissue | RIN at −70 °C | RIN at −80 °C | Mean difference |
|---|---|---|---|
| Whole blood RNA | 1.5, 1.6, 1.7 | 1.5, 1.6, 1.6 | 1.60 against 1.57 |
| Kidney | 5.9, 6.8, 5.8 | 7.0, 7.1, 7.3 | 6.17 against 7.13 |
| Lung | 5.6, 5.4, 5.6 | 2.3, 5.8, 6.0 | 5.53 against 4.70 |
The presenter concludes there were no significant differences. With three samples per group, the kidney gap of about one RIN unit fits that conclusion only if the within-group spread is large; the slides give no statistics and no paper was found. The lung mean is dragged down by a single −80 °C value of 2.3, and the blood RIN values of 1.5 to 1.7 are low enough that only the most expressed genes were analysed.
Other evidence: Edinburgh and VWR found no study comparing sample viability at different temperatures; Dutch Green Labs points to a long-term inventory study (Bousema et al., 2020) that was not read here; and a distributor notes that viable cell storage needs separate consideration.
Door openings, frost and room temperature
From a university guide: limit each opening to 45 seconds; allow 10 minutes of recovery per minute open; expect a rise of 1.2 to 4.7 °C after 60 seconds and 0.2 to 2.8 °C after 15; expect a swing of 5 to 8 °C on opening a −80 °C freezer against up to 3.1 °C at −60 °C; a freezer about 80 percent full with no inventory triples door-open time and raises energy about 60 percent; and drawers cut the temperature rise by about 65 percent.
Upkeep from the same guide: clear frost every 3 months, defrost above about 2 cm of ice for about 10 percent better efficiency, clean filters, evaporator coils and compressors every 3 months, keep 20 cm clear on top and 10 to 15 cm at the back and sides, hold the room at 15 to 18 °C for about 2 percent more energy per 1 °C above that, expect about 3 percent more energy per year of age, and up to 25 percent more with poor maintenance.
Ten years at 3 percent a year is 30 percent — the same size as the saving from −70 °C. A room at 25 °C instead of 18 °C is 14 percent more energy. Ice at 2 cm on 8.4 kWh/day is 0.84 kWh/day. A 60-second opening at 10 minutes of recovery per minute means about 10 minutes of recovery. A manufacturer case study reports that racking and a basic inventory cut door-open time by up to 67 percent and door-open energy by up to 70 percent, without publishing the underlying data.
Heat load and the building
Each 1 kWh saved at the freezer saves about 0.11 kWh of HVAC electricity, with a penalty of about 0.023 therms, according to a slide deck citing Paradise et al., 2019. On Edinburgh's 2.4 kWh/day saving the HVAC electricity saved is 0.26 kWh/day, so 2.66 kWh/day in total — 971 kWh a year against 876 at the freezer alone — with a penalty of 0.055 therms per day, or 20 therms a year.
Cost per freezer
At assumed tariffs of $0.10, $0.15 and $0.25 per kWh — round numbers, not quoted from a source — Edinburgh's 6.0 kWh/day is 2,190 kWh a year and its 8.4 kWh/day is 3,066. The 876 kWh difference is worth $87.60, $131.40 or $219.00 a year depending on the tariff. A 20 kWh/day freezer at $0.15 is $1,095 a year, and fifty freezers at the Edinburgh difference is 43,800 kWh a year.
Copenhagen's own estimate for a complete change across the university is 400 to 480 MWh a year, worth 85,000 to 105,000 EUR, or 0.6 percent of its 2017 energy budget — 212 to 219 EUR per MWh. Bath's figure of about £458 per freezer per year includes a UK tariff and is not comparable with the dollar rows.
Who has switched
CU Boulder reports about 75 of about 150 ULT freezers at −70 °C on its current page, and about half of about 175 on an earlier document. CRUK Manchester Institute moved 55 of 61 freezers from −80 °C to −70 °C. UCL Institute of Child Health has 46 of 67 at −70 °C, with changes starting in 2017. One Radboud group reports all 18 of its ULT freezers converted. Bath reports 24 freezers saving about £11,000 a year, or about £458 per freezer.
Manchester's share is 90 percent. The CU Boulder counts differ between pages, which is left unresolved.
Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 1600542033627 during review. CPS resolution does not verify the seller's technical claims, the setpoint range, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.
A −86 °C setpoint is a colder number than the standards test at, and the two figures that decide the running cost are missing: kWh per day at −75 °C, which is what ENERGY STAR measures, and the volume, which turns that into the kWh/day/ft³ the limits are written in. Ask for both, ask whether the setpoint is adjustable to −70 °C without losing the warranty, and ask for warm-up data or a CO2 backup option, since the difference between −70 and −80 °C in a power cut is three hours of rescue time. Sample type decides whether that matters: the published RIN comparison found no clear difference, and no peer-reviewed study was found that settles it.
What this article could not check
The ENERGY STAR test method text, which sets the ambient temperature and door procedure, was not available, nor was a peer-reviewed study comparing sample recovery at −70 and −80 °C across many sample types, nor the raw data behind the California 37 percent figure and the Graham et al. figure, nor the Bousema et al. study. No source explains the nine-hour and five-hour warm-up figures, and no test data on a −86 °C unit against the ENERGY STAR limits was found.