Selection GuideWorkshop & machinery15 min read

Laboratory Thermometers: Total or Partial Immersion, and Pt100 Classes That the Stem Correction Swamps

A 0.627 °C stem correction is 1.9 times the Class A tolerance of the sensor reading it. Below 200 °C the immersion, not the platinum, sets the uncertainty.

Updated November 18, 2026How Equipvia researches
A glass laboratory thermometer with a printed scale lying on a bench
The immersion line and the inscription on the reverse decide how the thermometer is meant to be read, and both are easy to lose in a catalogue. Photograph via Wikimedia Commons (CC BY-SA 3.0). View image source.

Disclosure: We may earn a commission if you purchase through links on this page, at no additional cost to you. Our selection criteria do not depend on commission; see our affiliate disclosure.

A mercury thermometer reading 90.15 °C with 60.15 degrees of its stem outside the bath needs a stem correction of +0.627 °C in the worked example published by a calibration laboratory. A Pt100 Class A sensor at 90.15 °C has a tolerance of ±0.330 °C under the IEC 60751 formula. The correction is therefore 1.9 times the Class A tolerance. Below, the immersion rules for liquid-in-glass thermometers, the IEC 60751 tolerance classes, the figures where sources disagree, and the verification intervals ASTM E77 gives.

Total, partial and complete immersion

ASTM E77 defines the two main groups. A partial-immersion thermometer is designed to indicate temperature correctly when the bulb and a specified part of the stem are exposed to the temperature being measured. A total-immersion thermometer is designed to indicate correctly when just the portion containing the liquid is exposed. E77 also defines a complete-immersion thermometer as one for which the entire thermometer is exposed, and notes that such a thermometer is not specified in ASTM documents.

The identification rules are practical rather than marked. A partial-immersion thermometer usually has an immersion line on the stem and an inscription on the reverse such as "76MM IMM". Some have no line: those with standard taper joints, and those with very short immersions such as melting point thermometers. A total-immersion thermometer usually has no line and no marking. If there is no inscription on the reverse, assume total immersion. Better makers inscribe TOTAL or TOTAL IMMERS.

Most total-immersion thermometers can also be used under complete immersion, for example inside a refrigerator, freezer or incubator. A total-immersion thermometer should be immersed to within about half an inch of the meniscus, which is 12.7 mm.

What a wrong immersion costs

The published example is a 24-inch total-immersion thermometer graduated from −1 to 201 °C, immersed only about 2 inches in a bath.

Quantity Value
Reading 190 °C
Error stated almost 5 °C, liquid hotter than indicated
Length 24 in = 609.6 mm
Immersion 2 in = 50.8 mm
Share of length immersed 8.3 percent
Error as share of reading 2.6 percent

ASTM E1 states that a thermometer used with the emergent stem at a different temperature than its calibration needs a calculated correction, with E77 giving the procedure. The correction is k × n × (T − t), where k is 0.00016 for mercury in Celsius, n is the number of degrees of the scale emergent from the bath, T is the observed reading, and t is the average temperature of the emergent column, found with a second total-immersion thermometer hung beside the first.

The worked example and its variations

Observed reading T: 90.15 °C. Emergent degrees n: 60.15. Emergent column temperature t: 25 °C. k for mercury in Celsius: 0.00016. Correction: 0.627 °C, corrected reading 90.777 °C. The same case with a red liquid at k = 0.001 gives 3.918 °C and 94.068 °C, a ratio of 6.25 between the two k values.

Variations on the same T and n with k = 0.00016:

  • t = 20 °C: correction 0.675 °C
  • t = 25 °C: correction 0.627 °C
  • t = 40 °C: correction 0.483 °C
  • n = 40.15 at t = 25 °C: correction 0.419 °C

At t = 40 °C the correction is 77 percent of the value at 25 °C, so a warm room or a heated bath top changes it by 0.144 °C between those two ambient conditions. With k = 0.001 the correction at t = 40 °C is 3.017 °C.

Two more cases make the scale of this clear. At T = 150 °C, n = 119 and t = 25 °C the correction is 2.38 °C; at t = 40 °C it is 2.09 °C. The first is 5.3 times the Class A tolerance of ±0.45 °C at 150 °C. At T = 90.15 °C the 0.627 °C correction is 6.3 times the Class AA tolerance at 0 °C and 1.9 times the Class A tolerance at that temperature. A correction 2 to 5 times larger than the sensor tolerance means the reading, not the sensor class, sets the uncertainty of the measurement.

Verification under ASTM E77

E77 covers visual and dimensional inspection, a pigment permanency test, a bulb stability test, and a scale accuracy test for liquid-in-glass thermometers specified in E1. It states that the scale readings of a thermometer change with time and use, and describes a reference point as a temperature at which a thermometer is checked for changes in bulb volume.

Ice-point practice: wait at least 3 minutes before reading, and successive readings taken at least 1 minute apart should agree within one tenth of a graduation. ASTM E563 is the practice for the ice-point bath. For direct-reading resistance thermometers used as secondary standards, E77 states a complete recalibration every 6 to 12 months depending on the temperatures of use, and ice points every 3 months. That is two to one recalibrations a year and four ice points a year.

The E77 text for liquid-in-glass thermometers used as secondary standards splits them into two groups: those for testing general-purpose total or partial-immersion thermometers, and those for testing special-use partial-immersion thermometers. E77 carries a warning that mercury is designated as hazardous by the EPA and many state agencies, and ASTM E2251 is the specification for liquid-in-glass thermometers with low-hazard precision liquids.

Pt100 tolerance classes under IEC 60751

A Pt100 has 100 Ω at 0 °C. Its temperature coefficient under IEC 60751 is 0.00385 Ω/Ω/°C, and an older US curve uses 0.003916. The tolerance in °C, with |t| the temperature without sign:

Class Tolerance, °C
AA ±(0.10 + 0.0017 |t|)
A ±(0.15 + 0.002 |t|)
B ±(0.30 + 0.005 |t|)
C ±(0.60 + 0.01 |t|)

Lapp, Rotronic, Pressure-Sense and an RS sensor sheet agree on these four formulas. WIKA states that the 2008 edition kept classes A and B and added AA and C: AA answered demand for more accurate thermometers, and C reflects the larger inaccuracy of film resistors above 500 °C.

Computed from the formulas:

t, °C AA A B C
−50 0.185 0.25 0.55 1.1
0 0.10 0.15 0.30 0.60
50 0.185 0.25 0.55 1.1
100 0.27 0.35 0.80 1.6
150 0.355 0.45 1.05 2.1
200 0.44 0.55 1.30 2.6
300 0.61 0.75 1.80 3.6
450 0.865 1.05 2.55 5.1

Cross-checks against the sources: Lapp lists Class AA at ±0.1 °C at 0 °C and ±0.27 °C at 100 °C, Class A at ±0.15 and ±0.35 °C; Rotronic gives Class B at 200 °C as ±1.3 K; the RS sheet lists Class A and B at 0, 100, 200, 300 and 400 °C as ±0.15 and ±0.3, ±0.35 and ±0.8, ±0.55 and ±1.3, ±0.75 and ±1.8, ±0.95 and ±2.3 °C. All match the table. RS also lists the resistance tolerance: Class A ±0.06 Ω at 0 °C and ±0.13 Ω at 100 °C, Class B ±0.12 Ω and ±0.30 Ω.

Ratios of Class B to Class A: 2.0 at 0 °C, 2.29 at 100 °C, 2.36 at 200 °C, and 2.43 at 450 °C. Class A to Class AA: 1.5 at 0 °C, 1.30 at 100 °C. Class C over Class B is 2.0 at 0, 100, 200 and 300 °C. The ratios approach fixed limits as temperature grows: 0.005 / 0.002 = 2.5 for B over A, 0.01 / 0.005 = 2.0 for C over B, and 0.002 / 0.0017 = 1.18 for A over AA. Rotronic also lists fractional classes at 0 °C: B 1/3 at ±0.1 K, B 1/5 at ±0.06 K and B 1/10 at ±0.03 K, which are those fractions of the Class B ±0.3 K.

Where the sources disagree on range

The classes have a validity range, and wire-wound and thin-film sensors differ.

  • Lapp, Class AA: −50 to +250 °C wire-wound, 0 to +150 °C thin-film.
  • Lapp, Class A: −100 to +450 °C wire-wound, −30 to +300 °C thin-film.
  • WIKA, Class B: −196 to +600 °C wire-wound, −50 to +500 °C thin-film.
  • RS sheet, 0.3 class: −196 to +660 °C wire-wound, −50 to +500 °C thin-film; 0.6 class: −196 to +660 °C wire-wound, −50 to +600 °C thin-film.

The wire-wound upper limit of Class B is 600 °C in the WIKA sheet and 660 °C in the RS sheet, a difference of 60 °C that the IEC text would settle and this article did not. WIKA also warns that a measuring resistor may be Class A while the finished thermometer has a different range, so the tolerance can change once the sensor is built into a probe — the class on the resistor is not automatically the class of the instrument.

Pt100 resistance and sensitivity

The RS sheet lists nominal resistance as 18.52 Ω at −200 °C, 60.26 Ω at −100 °C, 100.00 Ω at 0 °C, 138.51 Ω at 100 °C, 175.86 Ω at 200 °C, 212.05 Ω at 300 °C and 247.09 Ω at 400 °C. Mean sensitivity per 100 °C step: 0.3851 Ω/°C from 0 to 100 °C, 0.3735 from 100 to 200, 0.3619 from 200 to 300 and 0.3504 from 300 to 400, so the last is 91 percent of the first. From −100 to 0 °C it is 0.3974 Ω/°C. To resolve 0.01 °C at 100 to 200 °C a meter has to resolve 3.7 mΩ, and to resolve 0.001 °C it has to resolve 0.37 mΩ.

Tolerance in ohms from the same sheet, Class A and Class B: ±0.24 and ±0.56 Ω at −200 °C, ±0.14 and ±0.32 Ω at −100 °C, ±0.06 and ±0.12 Ω at 0 °C, ±0.13 and ±0.30 Ω at 100 °C, ±0.20 and ±0.48 Ω at 200 °C.

The RS sheet states that optimum accuracy is reached with a four-wire connection and defines a self-heating coefficient E = Δt / (R · I²), with Δt the indicated temperature minus the medium temperature. Pressure-Sense gives the Callendar–Van Dusen equation for IEC 60751: R(t) = R0(1 + At + Bt²) for 0 to +850 °C, with an added cubic term for −200 to 0 °C. Pt100 sensors in some humidity probes are Class B 1/3, as Rotronic notes for its HygroClip2 Advanced probes.

Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 1601251596217 during review. CPS resolution does not verify the seller's technical claims, the immersion type, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.

The listing answers the question this article spends its first half on: it states total immersion, a 0–200 °C range, and an organic liquid fill. The last of those is the one to notice. An organic fill has a coefficient around 0.001 rather than mercury's 0.00016, so the same 60 emergent degrees that cost 0.627 °C with mercury costs 3.918 °C here — a factor of 6.25 — and the thermometer is specified for immersion to within half an inch of the meniscus. Used that way it is fine; used half-dipped in a beaker it is not, and the error runs to degrees, not tenths.

Two things are still worth asking for. The listing shows a chart of several ranges rather than one certificate, so confirm which range the quoted price covers. And ask whether an ice-point certificate comes with it, since E77 asks for one every three months on a working standard, and the organic fill is the part that ages.

Choosing between the two families

Class B over Class A at 100 °C is 0.8 / 0.35 = 2.29. Class A at 300 °C is 0.15 + 0.002 × 300 = 0.75 °C, and at 450 °C it is 1.05 °C. Lapp's Class A range runs to 450 °C wire-wound and 300 °C thin-film, so a thin-film Class A at 400 °C is outside its own specification. Against that, the stem correction at 60 emergent degrees is 0.627 °C with mercury and 3.918 °C with red liquid, and E77 asks for recalibration every 6 to 12 months and an ice point every 3 months on a resistance thermometer used as a secondary standard.

The conclusion that follows is uncomfortable for anyone shopping on class alone: at 90 °C a Class A Pt100 and a Class AA Pt100 differ by 0.05 °C, while the immersion error on a liquid-in-glass thermometer in the same bath is 0.627 °C. Spending on the class before fixing the immersion buys precision that the installation throws away.

What this article could not check

The full text of IEC 60751:2008 and ASTM E1 was not available, so the ASTM E1 scale error limits by thermometer number are absent, as is an independent measurement of stem errors on a real bath and any manufacturer's published stem-correction table for partial-immersion thermometers used at the wrong depth. The worked examples come from a calibration laboratory, so they are a service provider's arithmetic rather than a standard's. The value of k for mercury could not be confirmed from the ASTM text itself, only from that example.