Selection GuideWorkshop & machinery16 min read

Laboratory Ovens: ASTM E145 Type IA, IB, IIA and IIB, the Uniformity Limits Worked Out at 105 and 150 °C, and the 27-Point DIN 12880 Test

The same 150 °C setpoint allows ±2.56 °C in a Type IA gravity oven and ±1.28 °C in a Type IIA forced-air oven — and no vendor figure can be compared with either until the test standard is named.

Updated November 20, 2026How Equipvia researches
A laboratory drying oven with its door closed on a bench
Uniformity is quoted as a percentage of the oven-to-room differential, so the same oven passes or fails depending on the room it stands in. Photograph via Wikimedia Commons (CC0). View image source.

Disclosure: We may earn a commission if you purchase through links on this page, at no additional cost to you. Our selection criteria do not depend on commission; see our affiliate disclosure.

ASTM E145 sorts general-purpose laboratory ovens into four types: IA and IB for gravity convection, IIA and IIB for forced ventilation. The uniformity limits, as read from a 1994 (2011) edition, are ±2, ±5, ±1 and ±2.5 percent of the difference between oven and room temperature. Vendors quote different numbers at 150 °C, measured to different methods, so the two sets of figures do not line up directly. Below, the limits are worked out at real setpoints, the vendor numbers are set beside them, and the disagreements are marked.

The four types

  • Scope: general-purpose air ovens with a chamber up to 0.6 m³ (25 ft³).
  • Gravity-convection ovens: from 20 °C above ambient to 200 °C.
  • Forced-ventilation ovens: from 20 °C above ambient to 500 °C.
  • Type IA: gravity convection, uniformity within ±2 percent of the oven-to-ambient differential.
  • Type IB: gravity convection, ±5 percent.
  • Type IIA: forced ventilation, ±1 percent.
  • Type IIB: forced ventilation, ±2.5 percent.

The plus-or-minus signs are lost in the excerpt text, which reads "62 %", "65 %" and so on; they are read here as ±2, ±5, ±1 and ±2.5, and the Table 1 values below confirm that reading. The standard does not cover safety for ovens used with combustible vapours or gases.

Table 1, as read from the 1994 (2011) edition

Characteristic Type IA Type IB Type IIA Type IIB
Deviation, differential 50 °C or less (max, °C) 1 2.5 0.5 1.25
Deviation, differential above 50 °C (% of differential) 2 5 1 2.5
Time constant (max) 600 s 720 s 480 s 660 s
Ventilation, minimum (air changes/hour) 10 10 50 50
Ventilation, maximum (air changes/hour) none none 200 200

The two deviation rows join without a step: at a 50 °C differential the percentage rows give 1.0, 2.5, 0.5 and 1.25 °C, the same as the first row. Table 1 from the 2019 edition was not available, so whether any value changed cannot be stated.

What the limits mean in degrees

With a 22 °C room, the limit is the percentage times (setpoint minus 22 °C). Below a 50 °C differential the absolute row applies.

Setpoint Differential Type IA Type IB Type IIA Type IIB
60 °C 38 °C ±1.00 °C ±2.50 °C ±0.50 °C ±1.25 °C
105 °C 83 °C ±1.66 °C ±4.15 °C ±0.83 °C ±2.08 °C
150 °C 128 °C ±2.56 °C ±6.40 °C ±1.28 °C ±3.20 °C
180 °C 158 °C ±3.16 °C ±7.90 °C ±1.58 °C ±3.95 °C

The best gravity limit is twice the best forced limit at every setpoint. At 105 °C — the usual glassware and moisture drying temperature — the Type IIA window is 1.66 °C wide and the Type IB window 8.30 °C wide, a ratio of 5.

Above 150 °C, for forced ventilation only, the limits are ±1.78 and ±4.45 °C at 200 °C, ±2.28 and ±5.70 at 250 °C, ±2.78 and ±6.95 at 300 °C, and ±4.78 and ±11.95 at 500 °C. Types IA and IB stop at 200 °C, where their limits are ±3.56 and ±8.90 °C.

Room temperature moves the limits, but not much: at 105 °C a Type IIA oven allows ±0.90 °C in a 15 °C room, ±0.83 in a 22 °C room and ±0.75 in a 30 °C room — 0.15 °C across a 15 K range.

How the uniformity test is run

Item ASTM E145 DIN 12880:2007-05
Points 9 27 above 50 L, 9 up to 50 L
Layout 8 corners at 5 cm from walls, 1 within 2.5 cm of centre 9 on each of 3 levels
Thermocouple iron or copper-constantan, 0.5 mm, junction up to 2 mm not given
Chamber state empty, shelves in place, vents open not given
Heat-up steady state, then record at least 24 h 2 h to steady state
Reading rate not given 6 per minute
End of heat-up not applicable centre stays inside the limit deviation (old DIN: 98 % of setpoint)
Example limit deviation not given ±1.5 K at 180 °C

The two sources conflict on ASTM. A trade summary says ASTM measures without preheating and puts four points on the top shelf, four on the bottom and one in the middle. The 1994 text says to bring the oven to temperature, wait for steady state and record for 24 hours, with thermocouples in the corners and centre. That is not resolved here: the standard's text was read, the trade summary is a summary.

Time constant and ventilation tests

The time constant test heats the oven to within 10 °C of its maximum operating temperature and holds at least an hour. A smooth brass cylinder about 9.5 mm in diameter and 57 mm long, with a differential thermocouple soldered on, is hung near the centre with the door open for one minute, then the differential is recorded every 10 seconds. The time constant is the time for the differential to fall to one tenth.

The ventilation test runs a sealed pass and a vented pass of at least 30 minutes each at 80 ± 2 °C above ambient, with ambient held within 0.2 °C, and uses N = 3590 (X − Y) / (V · D · ΔT), with the power difference in watts, volume in cm³, density in g/cm³ and ΔT in °C.

Worked on a 100 L chamber with D = 0.0012 g/cm³ and ΔT = 80 °C, the denominator is 9,600, so the power needed is 40 W for 15 air changes an hour, 26.7 W for the 10 that gravity ovens must reach, 133.7 W for the 50 that forced ovens must reach, and 534.8 W for the 200 maximum. Checking the formula against direct heating of the incoming air — 1.2 kg/m³, 1005 J/(kg·K), 80 K — gives 26.8, 133.9 and 535.6 W: agreement within 0.2 percent.

The time constant limits run 480 to 720 seconds, a spread of four minutes, with Type IIA 20 percent shorter than Type IA. Read as a first-order decay, which is an assumption and not something the excerpt states, dividing by ln 10 gives 208 s for IIA, 261 s for IA, 287 s for IIB and 313 s for IB.

Vendor numbers at 150 °C

From a major maker's general-protocol line, as listed by a distributor:

  • Gravity models: uniformity ±3.5 °C, stability ±0.4 °C at 150 °C.
  • Mechanical convection models: uniformity ±3.3 °C, stability ±0.3 °C.
  • Advanced protocol gravity: ±2.9 °C or better, stability ±0.3 °C.
  • One mechanical model: ±1.5 °C, stability ±0.2 °C.
  • The maker's own page: as good as ±1.3 °C at 150 °C.
  • 60 L models: 170 W gravity and 259 W mechanical to hold 150 °C.

Other makers: a 30 L forced-air oven is listed at not above ±2.5 percent of maximum temperature; a vacuum oven catalogue gives ±3.0 °C at 60 °C, ±9.0 °C at 120 °C and ±13.0 °C at 200 °C, with a separate line saying ±6 percent of setpoint, and states that its specifications follow DIN 12880; a third maker recommends forced-air ovens for multiple samples and higher moisture, and notes that many ASTM and AASHTO methods require forced draft.

Set against the E145 limits at 150 °C with a 22 °C room: ±3.5 °C is inside Type IB (±6.40) and outside Type IA (±2.56); ±3.3 °C misses Type IIB (±3.20) by 0.1 °C; ±2.9 °C is inside Type IIB; ±1.5 °C is inside Type IIB and outside Type IIA (±1.28); and ±1.3 °C misses Type IIA by 0.02 °C. The spread of the maker's own uniformity figures is 1.3 to 3.5 °C, a ratio of 2.7, and of the stability figures 0.2 to 0.4 °C, a ratio of 2.

These vendors test to DIN 12880 and not to E145, so the comparison is a rough guide only: a vendor ±3.3 °C figure does not show that the oven would pass Type IIB under E145.

Energy

From the 60 L figures, 170 W is 1,489 kWh a year at 24 hours a day and 259 W is 2,269 kWh — a difference of 780 kWh, a ratio of 1.52, or $223 against $340 a year at $0.15 per kWh. The two models are not described as having the same uniformity, so the extra 89 W is not a like-for-like price for tighter uniformity.

Ventilation in cubic metres and watts

A 100 L chamber takes 0.1 m³ per air change, so 10 air changes an hour is 1 m³/h, 0.59 cfm; 50 is 5 m³/h, 2.9 cfm; and 200 is 20 m³/h, 11.8 cfm. At the E145 size limit of 0.6 m³, 50 air changes an hour is 30 m³/h, which takes about 804 W of heating at ΔT 80 °C.

Percent specs against degree specs

A 300 °C forced-air oven quoted at not above 2.5 percent of maximum temperature would be ±7.5 °C if that means 2.5 percent of 300, while Type IIB at a 300 °C setpoint with a 22 °C room allows ±6.95 °C. The listing does not say what temperature the percentage applies to, so the match is uncertain.

One vacuum oven catalogue gives both forms: ±6 percent of setpoint, which is ±3.6 °C at 60 °C, ±7.2 at 120 and ±12.0 at 200; and ±3.0, ±9.0 and ±13.0 °C at the same three temperatures. The two forms differ by 0.6 °C at 60 °C, 1.8 °C at 120 °C and 1.0 °C at 200 °C, and the catalogue does not explain the difference. One other maker lists an oven at both ±1.5 °C and ±3.0 °C at 150 °C in two lines of the same document.

Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 1601909022069 during review. CPS resolution does not verify the seller's technical claims, the test standard, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.

A gravity oven is the right choice when the load is powders or when a fan would blow the sample around, and the wrong choice when the specification says ±1 percent. The number to ask for is the uniformity in degrees at the temperature actually used — 105 °C for glassware, 150 °C for general drying — with the test standard named, because a DIN 12880 figure and an E145 type are not interchangeable. Ask too whether the 42 L figure is the internal volume and how many measuring points were used.

Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 1601255466605 during review. CPS resolution does not verify the seller's technical claims, the test standard, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.

A forced-air oven has to reach 50 air changes an hour to qualify as Type II under E145, which on a 100 L chamber is 5 m³/h and about 134 W of heating at an 80 K rise — the fan is not the expensive part, the replacement air is. Ask for the air-change rate alongside the uniformity figure, because a fan that circulates without venting will hold a tight gradient while never clearing solvent vapour. Many ASTM and AASHTO methods require forced draft, so check the method before choosing gravity.

Reading a spec sheet

Numbers worth finding, with the figures from the sources above:

  • Test standard: DIN 12880 or ASTM E145. They are not interchangeable.
  • Number of measuring points: 9, or 27 above 50 L.
  • Uniformity in °C at the temperature you use, such as 105 °C or 150 °C.
  • Stability in °C: 0.2 to 0.4 °C in the maker's lines.
  • Empty or loaded chamber.
  • Watts to hold the setpoint: 170 and 259 W for 60 L, which is 2.83 and 4.32 W/L.
  • Chamber volume, so watts per litre can be worked out.

Operating windows from the E145 scope with a 22 °C room: gravity types run from 42 to 200 °C, a span of 158 K; forced-air types run from 42 to 500 °C, a span of 458 K, or 2.9 times the gravity span.

Standard ovens are not made for flammable solvents: the E145 scope excludes safety requirements for combustible vapours, which is a separate equipment class.

What this article could not check

Table 1 from ASTM E145-19, and any change from the 1994 (2011) values, was not available, nor was the DIN 12880:2007-05 limit deviation table beyond the ±1.5 K at 180 °C example, nor independent test data for any vendor figure, nor a door-open recovery time measured the same way on gravity and forced ovens, nor a source for what airflow real users need for glassware drying.