ASTM E287 gives a Class A tolerance of ±0.02 mL for a 10 mL buret, ±0.03 mL for 25 mL and ±0.05 mL for 50 mL, and says Class B is twice that: ±0.04, ±0.06 and ±0.10 mL. DWK sells a Class B burette to ISO 385 at ±0.03, ±0.04 and ±0.08 mL for the same three sizes, which is 1.5, 1.33 and 1.6 times the ASTM Class A figures and not 2.0. The smallest graduation differs as well: 0.05 mL on an ASTM 10 mL buret and 0.02 mL on the ISO ones.

Tolerance and graduation by size

The ASTM column is Table 1 of ASTM E287-02(2012) as printed on the standard's abstract and sample pages. The Class A tolerance there is ±0.02 mL at 10 mL, ±0.03 mL at 25 mL, ±0.05 mL at 50 mL and ±0.10 mL at 100 mL. The ISO/R 385:1964 recommendation, the first ISO text, lists Class A at the same four values. Its Class B rows come from a sample page that printed poorly, so the 25 mL Class B entries are a best reading.

Nominal, mL ASTM smallest division, mL ASTM A, ±mL ASTM B, ±mL ISO/R 385:1964 A ISO/R 385:1964 B
10 0.05 0.02 0.04 0.02 (div. 0.05) 0.05
25 0.1 0.03 0.06 0.03 (div. 0.05) 0.05 (div. 0.05), 0.1 (div. 0.1)
50 0.1 0.05 0.10 0.05 0.10
100 0.2 0.10 0.20 0.10 0.20

The 1964 text says the tolerance is the maximum permissible error at any point and also the maximum permissible difference between the errors at any two points. In neither class may the limit of error exceed the smallest scale division, which ISO 385-1:1984 repeats. At 10 mL with a 0.02 mL division, Class A at ±0.02 mL equals exactly one division. At 25 mL with 0.05 mL, ±0.03 mL is 0.60 of a division.

ASTM E287 also states that Class B is twice Class A, that burets are calibrated to deliver at 20 °C, and that no after-drainage period is allowed. Class B burets need not carry the class designation or a serial number. The standard covers 10, 25, 50 and 100 mL straight burets with single-bore or three-way stopcock, a 100 mL buret with a 50 mL intermediate bulb, and self-zeroing burets in the four sizes. Listings disagree on its status: one catalogue shows an E287-02(2019) edition, while two others mark the 2012 edition inactive or withdrawn.

Vendor figures that do not match the standards

Source Class 10 mL 25 mL 50 mL 100 mL
Hirschmann (Class AS, ISO 385) AS 0.02 0.03 0.05 –
Marienfeld (Class AS) AS 0.02 0.03 0.05 0.10
Growing Labs (ASTM E287, batch certified) A 0.02 0.03 0.05 –
DWK Duran (ISO 385) B 0.03 0.04 0.08 0.15

All values are ± mL. The DWK page marks the 100 mL size as non-DIN/ISO. Marienfeld states that Class AS tolerances are half of Class B under DIN ISO. Applied to its own figures, that implies Class B at ±0.04, ±0.06, ±0.10 and ±0.20 mL, which are the ASTM Class B values and not the DWK ones. The DWK Class B figure at 10 mL, ±0.03 mL, is 0.75 of ASTM Class B and 0.6 of the ISO/R 1964 figure of ±0.05 mL; at 50 mL it is 0.8 of both.

One Fisher Kimax automatic buret carries two class labels on the same page: the bullet reads Class B and the specification field reads Class A, with a tolerance of ±0.04 mL at 10 mL, which is the ASTM Class B value. Length 515 mm, graduation 0.05 mL.

Graduation differs by standard. Hirschmann, Marienfeld and DWK print 0.02 mL on the 10 mL, 0.05 mL on the 25 mL and 0.1 mL on the 50 mL. Growing Labs, citing ASTM E287, prints 0.05 mL, 0.1 mL and 0.1 mL. The ISO division is therefore 2.5 times finer at 10 mL and 2.0 times finer at 25 mL, and equal at 50 mL. BRAND adds that some burettes carry subdivision marks or a glass that DIN EN ISO 385 does not prescribe, and these are sold as meeting the Class B error limits without meeting the whole standard.

Delivery time and waiting time

ASTM Table 2 gives the minimum outflow time for Class A by graduated length: 30 s at 15 cm, 35 s at 20 cm, 40 s at 25 cm, 50 s at 30 cm, 60 s at 35 cm, 70 s at 40 cm, 80 s at 45 cm, 90 s at 50 cm, 105 s at 55 cm, 120 s at 60 cm, 140 s at 65 cm and 160 s at 70 cm. Table 3 gives Class B by capacity: 25 s for 10 mL, 30 s for 25 mL, 50 s for 50 mL and 90 s for 100 mL. Table 2 applies to straight burets only, not self-zeroing ones.

Mapping the ASTM scale lengths in Table 1 to Table 2 gives the Class A minimum outflow. A 10 or 25 mL buret with a 350 to 450 mm scale: 60 to 80 s. A 50 mL buret at 500 to 600 mm: 90 to 120 s. A 100 mL buret at 550 to 650 mm: 105 to 140 s. Hirschmann prints a flow time of 35 to 45 s for all three of its Class AS sizes, with a 30 s waiting time. That is below the ASTM Class A minimum for the same nominal sizes, but the scale lengths differ: ISO/R 1964 lists 25 to 35 cm for a 10 mL buret, which maps to 40 to 60 s in the ASTM table, and 50 to 60 cm for a 25 mL buret with a 0.05 mL division, which maps to 90 to 120 s.

ISO 385-1:1984 defines two types: type I burettes with no waiting time, in Class A and Class B (ISO 385-2), and type II burettes with a 30 s waiting time, in Class A only (ISO 385-3). BRAND sells BLAUBRAND with a 30 s waiting time and SILBERBRAND without one, and says the waiting time is not needed during a titration, since the dropwise addition at the end takes as long as the waiting time of a Class AS burette. Growing Labs lists a 30 s waiting time on its ASTM-based Class A burettes. ASTM E287 states that no after-drainage period is allowed, which sits against the 30 s of ISO type II. Neither full text was read.

Leakage limits

ISO 385-1:1984 limits the leakage of a closed stopcock to one half of the smallest scale division in 50 min for Class A and one full division in 50 min for Class B. The same limit applies to a stopcock tested without grease. The 1964 recommendation asks for a test of at least 30 min. In mL per hour: a 10 mL Class A buret with a 0.02 mL division may leak 0.012 mL per hour, a 25 mL buret with a 0.05 mL division 0.03 mL per hour, and a 50 mL buret with a 0.1 mL division 0.06 mL per hour. The Class B limits are 0.024, 0.06 and 0.12 mL per hour, a ratio of 2.0 at every size.

That limit is where stopcock material enters the purchase, because a glass plug needs grease to seal and grease is what a "tested without grease" figure is meant to exclude. A PTFE plug needs none, which is why the same class letter can carry different leakage risk depending on the plug in the box.

Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 1600969425781 during review. CPS resolution does not verify the seller's technical claims, the class, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.

Two lines decide whether this burette does the job. The stopcock is glass, so it needs grease, and ISO's leakage limit is written for a stopcock tested without grease; ask for the leak test result in mL per 50 minutes at the size you buy. And "Class A" covers ±0.02 mL at 10 mL under ASTM E287 with a 0.05 mL division, or ±0.02 mL under ISO 385 with a 0.02 mL division, so ask which standard was used and what the smallest division is. A grade A mark printed without a standard behind it is not the same claim as a certificate.

Relative error and the 30 percent rule

USP general chapter <31> states that a buret should be of a size such that the titrant volume is not less than 30 percent of the nominal volume. It also says to use Class A apparatus unless a monograph says otherwise, that plastic volumetric apparatus accepted tolerances equal Class B glass, and that readings on 25 and 50 mL burets are estimated to the nearest 0.01 mL and on 5 and 10 mL burets to the nearest 0.005 mL. Its buret table lists a 10 mL micro type with a 0.02 mL division and ±0.02 mL, a 25 mL at 0.1 mL and ±0.03 mL, and a 50 mL at 0.1 mL and ±0.05 mL.

Relative error of the Class A tolerance at three delivered volumes:

Nominal, mL Tolerance, ±mL At full volume At 50 % At 30 %
10 0.02 0.20 % 0.40 % 0.67 %
25 0.03 0.12 % 0.24 % 0.40 %
50 0.05 0.10 % 0.20 % 0.33 %
100 0.10 0.10 % 0.20 % 0.33 %

A reading estimated to 0.01 mL is 20 percent of the 50 mL tolerance of 0.05 mL. Two readings, start and end, each uncertain by 0.01 mL, combine to 0.014 mL by root-sum-square or 0.020 mL in the worst case, which is 47 percent or 67 percent of the 25 mL tolerance of 0.03 mL. At 30 percent of a 25 mL buret, a 7.5 mL delivery carrying 0.014 mL of reading uncertainty is 0.19 percent before the 0.40 percent tolerance is considered.

The 30 percent rule is the reason a bench needs three sizes rather than one large buret. Delivering 7.5 mL from a 50 mL buret sits at 15 percent of nominal, below the USP floor, and costs 0.67 percent in tolerance terms against 0.33 percent at 30 percent of a 25 mL buret.

Scale length, divisions and bore

ASTM Table 1 allows a scale length of 350 to 450 mm for the 10 mL and 25 mL burets, 500 to 600 mm for 50 mL and 550 to 650 mm for 100 mL. The number of divisions is capacity over division: 200 for a 10 mL buret at 0.05 mL, 250 for 25 mL at 0.1 mL, 500 for 50 mL at 0.1 mL and 500 for 100 mL at 0.2 mL. That is 1.75 to 2.25 mm per division at 10 mL, 1.40 to 1.80 mm at 25 mL, 1.00 to 1.20 mm at 50 mL and 1.10 to 1.30 mm at 100 mL. The 1964 ISO recommendation lists 250 to 350 mm for 10 mL at 0.05 mL, which is 1.25 to 1.75 mm per division, and 500 to 600 mm for 25 mL at 0.05 mL, which is 1.00 to 1.20 mm per division.

A rough bore estimate divides capacity by scale length and ignores the ungraduated parts. A 10 mL buret over 400 mm has a cross-section of 0.25 cm², an inside diameter near 5.6 mm. A 25 mL buret over 400 mm has 0.63 cm², near 8.9 mm. A 50 mL buret over 550 mm has 0.91 cm², near 10.8 mm. The 50 mL Class A tolerance of 0.05 mL is a column length of 0.55 mm at that bore, a little over half of a 1.0 mm division.

Marienfeld sells a clear glass Class AS buret made from precision tubing, calibrated at 5 points and certified to DE-M with a batch certificate. Its 10 mL entry reads 10 : 0,02 ml, which is the capacity and the division, and ±0,02 ml, which is the tolerance. The Kimax 10 mL automatic buret is 515 mm long with a 0.05 mL graduation and a black ceramic enamel scale, and its three-way PTFE-plug stopcock refills through the inlet tube and adjusts to zero automatically.

Stopcock material and glass

Marienfeld sells the same Class AS tolerances with a glass plug and with a PTFE plug, at ±0.02, ±0.03, ±0.05 and ±0.10 mL, so the stopcock material does not change the class tolerance in its listing. Hirschmann lists a PTFE spindle valve, Growing Labs a PTFE screw-thread stopcock, and the Kimax automatic buret a No. 2 PTFE-plug three-way stopcock. ASTM E287 refers to ASTM E675 for interchangeable taper-ground stopcocks and stoppers. Hirschmann sells clear glass with a Schellbach stripe, and Marienfeld sells clear, Schellbach and amber glass with white graduation. DWK lists its Duran buret as borosilicate glass 3.3, calibrated to deliver at 20 °C.

The practical argument for a PTFE plug is not accuracy but maintenance: a glass plug that seizes, or that is greased unevenly, is the usual reason a buret is taken out of service, and neither standard read here gives a drift figure for either material. Hirschmann's and Marienfeld's listings put the same tolerance on both, which suggests the choice is about the solvent and the operator rather than the class.

What this article could not check

The full tolerance table of the current ISO 385 edition was not found, nor the ISO 385 Class A figures at 10 mL with a 0.02 mL division beyond vendor listings, nor a measured comparison of glass and PTFE stopcocks for leakage or drift. The relative errors above are tolerances, not measured errors. The ISO 385-1:1984 limits table was not legible in the sample.