15 mL vs 50 mL Conical Tubes: RCF Ratings, PP vs PS, and Why Catalogues Disagree
The same Falcon 15 mL tube is rated 12,000 × g in polypropylene and 1,800 × g in polystyrene. They sit on one distributor drop-down, one word apart.

Disclosure: We may earn a commission if you purchase through links on this page, at no additional cost to you. Our selection criteria do not depend on commission; see our affiliate disclosure.
Corning sells its 15 mL Falcon conical tube in polypropylene, rated 12,000 × g, and in polystyrene, rated 1,800 × g. Both come with blue dome-seal caps. TPP's 50 mL polypropylene tube is listed at 15,500 × g by one distributor and at 9,400 × g by another. Size alone tells you nothing about the rating.
RPM is the setting, RCF is the load
The dial says rpm. The tube box says × g. Rotor radius links them, so 4,000 rpm in one rotor is a different force in another. The common conversion is RCF = 1.118 × 10⁻⁵ × r × rpm², with r in centimetres, usually from the axis to the tube bottom.
Two worked examples. A swinging-bucket rotor, 15 cm radius, 4,000 rpm: about 2,680 × g. A fixed-angle rotor, 10 cm radius, 10,000 rpm: about 11,180 × g. Almost any plastic tube survives the first. A polystyrene tube will not survive the second, and some polypropylene tubes are close to their limit.
Thermo's application note on conical tubes says each tube has a maker-set speed rating, and that exceeding it can lose the sample and damage the rotor. It adds that fit matters. Load pressed straight onto the tube wall leaves stress lines, bulges or cracks; a tube that seats well hands the load to the rotor. Bangor University's safety sheet makes the same point as a pre-run check: look for cracks or crazing, and confirm the tube suits the rotor.
What the catalogues list
| Tube (nominal size) | Material | Listed maximum RCF | Source type |
|---|---|---|---|
| Falcon 15 mL | Polypropylene | 12,000 × g | Distributor, quoting the maker |
| Falcon 15 mL | Polystyrene | 1,800 × g | Distributor, quoting the maker |
| Falcon 50 mL | Polypropylene | 16,000 × g | Maker's catalogue |
| TPP 50 mL | Polypropylene | 15,500 × g | Distributor |
| TPP 50 mL | Polypropylene | 9,400 × g | Second distributor |
| TPP 50 mL | Polystyrene | 1,700 × g | Both distributors |
| Fisherbrand Easy Reader 50 mL | Polypropylene | 15,000 × g | Distributor |
| BRAND polypropylene | Polypropylene | 3,000 × g | Distributor |
| Labcon MetalFree 15 mL | Polypropylene | 17,000 × g | Distributor |
| Generic 50 mL | PP / PS | 3,000 / 1,000 × g | Distributor |
Every figure in the table is a seller's. Polystyrene appears at 1,000 (a generic 50 mL), 1,700 (TPP) and 1,800 (Falcon 15 mL). Polypropylene appears at 3,000 (BRAND, and the generic tube), 12,000 (Falcon 15 mL), 15,000 (Fisherbrand Easy Reader), 15,500 or 9,400 (TPP), 16,000 (Falcon 50 mL) and 17,000 (Labcon MetalFree 15 mL). A tube sold as plain polypropylene can therefore sit anywhere over a fivefold range.
Corning's figure comes from a fully supported rotor with room-temperature water. TPP's depends on adapter and temperature. None of the listings gives a rating for a frozen tube, a thick gradient or a half-filled tube.
Where the numbers conflict
TPP 50 mL polypropylene: one distributor prints 15,500 × g. Another prints 9,400 × g under what is word-for-word the maker's description. Both keep the footnote about adapter and temperature. Whether that is two catalogue editions, two tube lines or a typo cannot be settled without TPP's own datasheet, which the listings do not reproduce.
Falcon 15 mL: Corning lists the polystyrene and polypropylene tubes as separate catalogue numbers, 500 per case each, with names that differ by one word. Fisher's ordering page puts both on one drop-down, 12,000 or 1,800 × g. Reorder from an old purchase record, pick the wrong line, and the rating falls about sevenfold.
Outside diameter matters for rotor fit, and it conflicts too. One Fisher listing gives the Falcon 50 mL as 28 mm wide and 115 mm long. Another gives 30 mm. McMaster-Carr gives 30 mm for its 50 mL polypropylene tube and 17 mm for the 15 mL. Kimble's glass 50 mL tube is 29 mm. A 2 mm spread separates a snug fit from a rattle, or from a tube that will not go in at all. Take a tube that already fits your rotor and measure it against the one you plan to order.
Glass shows a smaller version of the same pattern. Two Kimble borosilicate 50 mL conical tubes carry listed maximum RCFs of 2,980 × g and 2,000 × g. They differ in length and bottom shape, so this is two products rather than one error.
Polypropylene or polystyrene
Polystyrene is clear, cheap and rigid. Polypropylene is slightly hazy, tougher and tolerant of a wider range of chemicals and temperatures. The sellers' spec sheets line up on this.
TPP gives −196 to +121 °C for its polypropylene tubes and −40 to +70 °C for polystyrene. A distributor description says polypropylene tubes can be autoclaved at 121 °C and resist acids, solvents and alkalis at room temperature, while polystyrene tubes tolerate mild bases and weak acids but not organic solvents or aromatic or chlorinated hydrocarbons, and cannot be autoclaved. A Nalgene guideline separately says not to autoclave polystyrene.
TPP's own chemical resistance table, which rates continuous exposure at 20 °C, gives acetone a good rating for polypropylene and a poor one for polystyrene, and gives chloroform an uncertain rating for polypropylene and a poor one for polystyrene. Carbon tetrachloride and chlorobenzene are poor for polypropylene as well. The same table warns that it is guidance and not an absolute value. Thermo's sample-storage compatibility guide makes the same caveat and says to test under your own conditions.
Cold storage follows the same split. Corning describes its polypropylene tubes from 15 to 225 mL as suitable for long-term storage at −80 °C. TPP's range for polypropylene runs down to −196 °C, while its polystyrene stops at −40 °C. A polystyrene tube that goes from the freezer into a rotor is the case worth worrying about, since cold plastic is stiffer and the rating assumes room temperature.
Caps vary too. Corning's Falcon 15 mL tubes use a dome-seal screw cap and the 50 mL tubes a double-threaded cap that the maker says reduces cross-threading. Fisherbrand sells plug-seal and flat-top versions of the same 50 mL tube. A cap that does not match the tube can leak long before the tube itself reaches its rating.
A working rule, which is judgment: if the tube will touch an organic solvent at any point in the protocol, buy polypropylene and check the specific solvent in the maker's table. If the sample is aqueous, the spin is gentle and you need to see the pellet clearly, polystyrene is fine, and it is the less expensive tube. Never spin polystyrene above the rating on its own box, and never assume a tube is polypropylene because it looks like the polypropylene one you used last week.
How full the tube should be
A trade-journal note on centrifuge consumables gives a general rule that a tube or bottle should be filled at least 75 per cent, and warns that a vessel less than half full can suffer high wall stress and fail. It also says some tubes, such as ultracentrifuge tubes, need to be completely full. Thermo's page for its polysulfone Oak Ridge tubes asks for at least 80 per cent fill for proper performance.
For a conical tube the practical consequence is awkward. A 50 mL tube holding 8 mL of a small sample falls well below those fractions. Two options follow. Use a 15 mL tube for the same sample, which is the better answer when the rotor accepts both sizes. Or, where the protocol allows, top up with buffer. "It has always worked" is not a reason to run a small sample at the top of the rated speed in a large tube.
The second fill issue runs the other way. A tube filled to the shoulder and spun in a fixed-angle rotor can leak from the cap, because the angled tube pushes liquid toward the opening. University of Connecticut guidance on fixed-angle rotors says never to overfill tubes, and to work out the maximum volume for the tube's angle, because liquid can leak out under the rotor lid. If you do not know the limit for your rotor, ask the person who maintains the instrument.
Equipvia may earn a commission from qualifying purchases through this link. The link resolved to Product ID 62458154457 during review. CPS resolution does not verify the seller's technical claims, the RCF rating, affiliate eligibility, or commission attribution. Check the actual selected variant and current terms before ordering.
Two details on this listing are worth reading against the sections above. It offers all three common sizes, 10, 15 and 50 mL, which is useful when the right answer is a smaller tube rather than topping up a large one, and it quotes an allowable error of 0.1 mm. That is a dimensional figure, not a volumetric tolerance, so it says nothing about how close the moulded graduations are to the volume they mark, which is the gap discussed later on this page. Ask for the maximum RCF with its rotor and temperature conditions, the sterility assurance level if the work needs one, and the nuclease certification if the tube will hold nucleic acids.
Biology samples: containment and the wait
When the sample is infectious, the tube is only the first barrier. A University of New Mexico biosafety incident summary describes screw-cap tubes of virus-containing material spun in a swinging-bucket rotor that had no safety cups. A tube seam cracked, and a swab of the chamber liner found about 4,000 plaque-forming units afterwards. The lessons listed include confirming that tubes are rated for the g-force, noting that polystyrene tubes can be mistaken for polycarbonate ones that tolerate more force, and using gasketed safety cups or sealed rotors for pathogens.
The waiting period afterwards is where sources disagree. The Iowa State biosafety manual says to open tubes in a biosafety cabinet, and where one is not available, to allow at least 10 minutes of settling. A trade-magazine safety article says to wait at least 30 minutes. The University of New Mexico report states that a 20-minute wait did not remove the hazard, since very small aerosol particles can stay airborne for more than an hour. These cannot all be right for the same situation. For infectious material, follow your institution's biosafety officer, not a number in a catalogue.
Stanford's biosafety-level table lists centrifuge safety cups or sealed rotors as the containment for biohazardous material at the higher levels, and the University of Tennessee guidance says to use sealed cups, buckets or O-ring rotors where available and to fill and open tubes in a biosafety cabinet.
Graduations are a guide, not a measurement
Printed marks on the side of a plastic tube are convenient, and sellers describe them in confident terms such as "accurate readings". No published volumetric tolerance for the graduations on a disposable polypropylene or polystyrene conical tube turned up.
The spacing also varies. Fisherbrand's Easy Reader 50 mL tube is marked every 5 mL from 5 to 50, and its 15 mL tube every 0.5 mL from 2 to 14.5 mL. A hardware catalogue lists 2.5 mL steps on a 50 mL polypropylene tube and 0.5 mL on the 15 mL. TPP says its white marks continue to the bottom of the cone so that small volumes can be read, and that the 15 mL scale starts at 0.1 mL.
Glass shows what a stated tolerance looks like. Kimble's 50 mL glass conical tubes, which the maker says are made to specifications in ASTM D2158 and are calibrated to contain, list a tolerance of ±0.075 mL up to 1 mL, ±0.30 mL between 2 and 5 mL, and ±1.0 mL above 10 mL. Even there, the error at the 50 mL mark is a full millilitre.
Use the graduations to see roughly how much is in the tube and to balance tubes against each other by eye. Use a pipette, a cylinder or a balance for any volume that matters to the result.
What to specify when you reorder
Write down five things and keep them with the purchase record: the nominal volume, the material, the rated RCF with its test conditions, the outside diameter if your rotor is tight, and the cap style (dome seal, plug seal or flat). Fisher's buying guide suggests skirted, free-standing versions when the bottom is conical, so tubes stay upright on the bench. Corning sells its 50 mL Falcon tube in bags of 25 or racked, 500 to the case, under different catalogue numbers. Two other details save trouble. Check that sterile or non-sterile matches the work, and that the pack is racked or bulk depending on how you handle them.
What this article could not check
None of the makers' datasheets for RCF test conditions were read; every rating above is a retailer's or maker's catalogue line, and the conflicting TPP figures remain unresolved. No independent test of centrifuge tube failure speeds was found, and no published tolerance for graduation marks on disposable plastic tubes. The biosafety waiting times come from institutional documents that were not compared with the full BMBL text. Before running a valuable or hazardous sample near the top of a tube's rating, confirm the figure with the maker's technical service, and consider a trial run with water.